    摘要:

    为进一步压缩检测时长、实现进出口食品快速通关,研制阵列装置用于多毒素同步快速检测。设计并以3D打印技术构建了卧式和立式2种适配于胶体金侧流免疫分析(LFIA)试纸的阵列装置,并对装置的性能进行评价。以阵列装置同时集成黄曲霉毒素B1(aflatoxin B1,AFB1)、玉米赤霉烯酮(zearalenone,ZEN)、赭曲霉毒素A(ochratoxin A,OTA)以及脱氧雪腐镰刀菌烯醇(deoxynivalenol,DON)等4种真菌毒素的LFIA试纸并应用于玉米、玉米面和麦仁等食品的检测。试验结果显示:4种毒素的检测限分别达0.031、0.190、0.780、0.220 ng/mL,回收率为74.42%~112.40%,结果无交叉反应,包括前处理时间在内单一样品的四指标同步检测可在40 min内完成。

    Abstract:

    An array device for synchronously and rapidly detecting multiple mycotoxins was developed to further reduce the time of detection and achieve rapid customs clearance of import and export food. Due to the interaction between probe and analyte,it is difficult to quantify the existing scheme of fixing multiple detection lines on a single test paper,and single quality control line cannot guarantee the effectiveness of multiple channels,and the real “synchronous” detection is not achieved. In this article,two kinds of horizontal and vertical arrays adapted to colloidal gold lateral flow immunoassay (LFIA) strips were designed and constructed by 3D printing technology,with four channels synchronous detection capability. The LFIA strip of aflatoxin B1 (AFB1),zearalenone (ZEN),ochratoxin A (OTA) and deoxynienol (DON) was integrated into the array and used to detect corn,corn meal and wheat kernel. Results showed that the detection limit (LOD) of the four toxins was 0.031 ng/mL,0.19 ng/mL,0.78 ng/mL,0.22 ng/mL,respectively,with recoveries ranging from 74.42% to 112.40%. There was no cross-reaction. The synchronous detection of the four indicators of a single sample,including the pretreatment time,was completed within 40 minutes. The array device constructed will provide a potential equipment support for synchronous and rapid detection of multiple mycotoxins.